$0.00
Dapttin® TC is a double activated aPTT reagent with sulfatides and silica as surface activators, together with an optimal mixture of highly purified phospholipids. The double activation of the contact phase provides high sensitivity to heparin over the complete measurement range, lupus sensitivity, sensitivity for clotting factor defects and a short incubation time. The reagent is optically clear and suitable for both optical and conventional coagulometers.
Dapttin® TC (double-activated aPTT) is a standardized aPTT-(activated partial thromboplastin time) reagent consisting of slilica and sulphatide as surface activators and a blend of highly purified phospholipids. Response to all factors and inhibitors is optimal.
(not supplied with the kit)
** or any other package sizes, special Ceveron® alpha or Technoclot® Control and Calibration reagents of Technoclone.
Plasma separation: Mix 9 parts of venous blood and 1 part of Sodium Citrate Solution (0.11 mol/l) and centrifuge for 15 minutes at a RCF of at least 2500 g (corresponding to DIN 58905). Store the plasma at room temperature and use within 4 hours. Stability at -20°C: 1 month.
Technoclone provides application sheets for Ceveron® alpha. The application sheets contain analyzer/assay specific handling and performance information which may differ from that provided in this instruction for use. In this case the information contained in the application sheets supersedes the information on the package insert on this product page. Please consult the instruction manual of the Ceveron® alpha.